134
Chapter 2
Front-Panel Key Reference
Source
F
ront-
P
ane
l K
e
y Ref
e
rence
Source
Accesses the tracking generator key functions (Options 1DN or 1DQ
only). Without Option 1DN or 1DQ, pressing
Source
will cause the error
message
Option not installed
to appear on the analyzer display.
CAUTION
If
Auto Align
is on, the 3 GHz tracking generator will be momentarily
retuned to approximately 1.557 GHz between most sweeps. Some
devices under test (for example, amplifiers with AGC) may be
susceptible to damage due to this momentary retuning. To avoid this
momentary retuning, turn the auto align off by pressing
System
,
Alignments
,
Auto Align
,
Off
. Refer to the Specifications Guide for your
instrument to learn more information about using the analyzer with
Auto Align set to off.
NOTE
The output frequency of the 3 GHz tracking generator may not exactly
match the input frequency of the spectrum analyzer. This is probably
not a problem when using wide resolution bandwidths. However,
narrowing the RBW may cause a drop in the displayed signal amplitude
as the frequency offset puts the signal out of the center of the RBW
filter. For narrower RBWs you may need to do a tracking peak to adjust
the tracking generator frequency. See the
Tracking Peak
and
Man
Tracking Adjust
.
Amplitude
On Off
Activates (
On
) or deactivates (
Off
) the output power of the tracking
generator. The power level can then be adjusted using the numeric
keypad, step keys, or knob. Pressing any digit, 0 through 9, on the
numeric keypad brings up the selected terminator menu. See the
specifications chapter in the calibration guide for the available output
power for your tracking generator.
Key Access:
Source
CAUTION
Power-level sensitive devices connected to the tracking generator
output may be accidentally damaged. This is because the actual source
amplitude will be greater than the amplitude indicated on the analyzer,
when the power sweep function is being used.
The source amplitude value is used as the starting amplitude for a
power sweep. The ending amplitude for a power sweep is higher then
the source amplitude value and it may actually be set to values beyond
specified output levels. Therefore, exercise caution when connecting a
power-level sensitive device to the tracking generator output.
Summary of Contents for ESA-E E4401B
Page 4: ...4 ...
Page 35: ...35 Using This Document 1 Using This Document ...
Page 38: ...38 Chapter1 Using This Document What is in This Book Using This Document ...
Page 196: ...196 Chapter3 Programming Fundamentals Overview of RS 232 Option 1AX Programming Fundamentals ...
Page 393: ...Chapter 6 393 Menu Maps Alpha Editor Menu Menu Maps Alpha Editor Menu ...
Page 394: ...394 Chapter6 Menu Maps AMPLITUDE Y Scale Menu Menu Maps AMPLITUDE Y Scale Menu ...
Page 396: ...396 Chapter6 Menu Maps Auto Couple Menu Menu Maps Auto Couple Menu ...
Page 397: ...Chapter 6 397 Menu Maps BW Avg Menu Menu Maps BW Avg Menu ...
Page 398: ...398 Chapter6 Menu Maps Det Demod Menu Menu Maps Det Demod Menu ...
Page 399: ...Chapter 6 399 Menu Maps Display Menu Menu Maps Display Menu ...
Page 400: ...400 Chapter6 Menu Maps File Menus 1 of 3 Menu Maps File Menus 1 of 3 ...
Page 401: ...Chapter 6 401 Menu Maps File Menus 2 of 3 Menu Maps File Menus 2 of 3 ...
Page 402: ...402 Chapter6 Menu Maps File Menus 3 of 3 Menu Maps File Menus 3 of 3 ...
Page 403: ...Chapter 6 403 Menu Maps Freq Count Marker Menu Menu Maps Freq Count Marker Menu ...
Page 404: ...404 Chapter6 Menu Maps FREQUENCY Channel Menu Menu Maps FREQUENCY Channel Menu ...
Page 405: ...Chapter 6 405 Menu Maps Input Output Menu Menu Maps Input Output Menu ...
Page 406: ...406 Chapter6 Menu Maps Marker Menu Menu Maps Marker Menu ...
Page 407: ...Chapter 6 407 Menu Maps Marker Menu Menu Maps Marker Menu ...
Page 408: ...408 Chapter6 Menu Maps Peak Search Menu Menu Maps Peak Search Menu ...
Page 409: ...Chapter 6 409 Menu Maps Preset Menu Menu Maps Preset Menu ...
Page 410: ...410 Chapter6 Menu Maps Print Setup Menu Menu Maps Print Setup Menu ...
Page 411: ...Chapter 6 411 Menu Maps Source Menu Menu Maps Source Menu ...
Page 412: ...412 Chapter6 Menu Maps SPAN X Scale Menu Menu Maps SPAN X Scale Menu ...
Page 413: ...Chapter 6 413 Menu Maps Sweep Menu Menu Maps Sweep Menu ...
Page 414: ...414 Chapter6 Menu Maps System Menu Menu Maps System Menu ...
Page 415: ...Chapter 6 415 Menu Maps Trace View Menu Menu Maps Trace View Menu ...
Page 416: ...416 Chapter6 Menu Maps Trig Menu Menu Maps Trig Menu ...
Page 426: ...426 Index Index Index ...