Installation Note E8356-90052
17
Step 6.
Perform Source and Receiver Calibration Adjustments
The following adjustments must be made due to the change in the full frequency range of the
analyzer.
Source Calibration Adjustment
The source calibration is used to adjust your network analyzer for a flat source power across
its full frequency range. The differences between Port 1 and Port 2 are negligible; therefore, a
Port 2 adjustment is not required.
1. Connect the equipment as shown in
. Connect a GPIB cable between the network
analyzer and the power meter.
Figure 7
Setup for the Source Calibration Adjustment
2. On the
System
menu, point to
Service
,
Adjustments
, then click
Source Calibration
.
3. Ensure the GPIB settings are correct.
4. Click
Calibrate
, then follow the instructions as they are displayed.