518
Chapter 15
Measurement Examples
Measuring Parameters with Cable
Measuring Parameters with Cable
This section introduces an example of how to detect the location of a mismatch that occurs
in a cable by using the time domain function.
Overview of evaluation procedure
In this example, a DUT is evaluated according to the steps shown in Table 15-8.
Table 15-8
Evaluation Procedure for Deviation from Linear Phase
Step
Description
“1. Setting the Measurement
Conditions” on page 518
Set the measurement conditions.
“2. Executing Calibration” on
page 518
Execute calibration.
“3. Connecting the DUT” on page 519
Connect the DUT.
Execute auto scale.
“5. Setting the Time Domain
Function” on page 520
Set the time domain function.
1. Setting the Measurement Conditions
Follow these steps to set the measurement conditions:
Setting Description
Key Operation
Presetting
-
OK
Stop frequency: 3 GHz
Number of points: 201
-
Points
-
Specifying the low-pass mode sweep condition
-
Transform
-
Set Freq Low Pass
Measurement parameter: S11
-
S11
2. Executing Calibration
By following “1-Port Calibration (reflection test)” on page 113, execute 1-port calibration
on port 1.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...