![Agilent Technologies E5070B User Manual Download Page 526](http://html.mh-extra.com/html/agilent-technologies/e5070b/e5070b_user-manual_2868065526.webp)
520
Chapter 15
Measurement Examples
Measuring Parameters with Cable
5. Setting the Time Domain Function
Set the conversion function to display the response in time the domain. If you enable this
setting, the response in time domain is displayed as shown in Figure 15-34. A peak
indicating a small mismatch appears at the location of the connector.
Setting Description
Key Operation
Data format: real
-
Real
Setting the transformation type to low-pass
impulse
-
Transform - Lowpass Imp.
Setting the window type to maximum.
Window
-
Maximum
Setting the display range: from 0 s to 10 ns
Start
-
Stop
-
Enabling the transformation function
Transform
(set to
ON
)
Executing auto scale
-
Auto Scale
Figure 15-34
Response in time domain
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...