Page 1: ...0 90660 Printed in USA August 2007 Agilent Technologies ESG Vector Signal Generator E4423B E4424B E4425B E4426B E4434B E4435B E4436B and E4437B Semi Rigid Cable Replacement Kits Kit Part Numbers E4400...

Page 2: ...to this material including but not limited to the implied warranties of merchantability and fitness for a particular purpose Agilent Technologies shall not be liable for errors contained herein or fo...

Page 3: ...daughterboard card cage boards 4 Remove the old semi rigid cables 5 Install the new semi rigid cables 6 Replace daughterboard card cage boards 7 Replace inner covers 8 Verify the functionality of the...

Page 4: ...660 Table 2 Replacement Cable Kit for W43 E4400 60751 Item Quantity Description Part Number 1 2 Retainer Ring 0510 1643 2 2 Retainer Cable E4400 40002 3 1 Cable out to Synth E4423 20026 4 1 Installati...

Page 5: ...NING Before you disassemble the instrument turn the power switch off and unplug the signal generator Failure to unplug the signal generator can result in personal injury CAUTION Electrostatic discharg...

Page 6: ...utes 2 Run the instrument self test by pressing Utility Instrument Info Help Mode Self Test Run Complete Self Test Upon completion a summary of the self test will be displayed Use the service guide to...

Page 7: ...feet 2 4 Remove the four bottom feet 5 and 6 from the cover by pushing and pulling the tab 5 Slide the outer cover 4 off the frame Removing the Inner Top Cover Refer to Figure 1 1 Using a T 10 driver...

Page 8: ...8 Figure 1 Outer and Inner Cover Removal...

Page 9: ...oard A11 Reference Board A23 Sampler Board A24 Frac N Board 1 Remove the right side cover by removing the four screws that attach it to the instrument chassis 2 Simultaneously lift the left and right...

Page 10: ...tor with Option UNB refer to or Figure 4 1 Refer to Figure 5 Remove the snap ring 1 from the cable retainer 2 of the cable you want to remove The cable is now disconnected from the instrument chassis...

Page 11: ...11 Figure 4 W42 W43 and W44 Semi Rigid Cable Locations for ESG AP and ESG DP Series Signal Generators with Option UNB...

Page 12: ...le 4 Verify the cable retainer is correctly attached by listening for a click as you slide the cable retainer over the end of the cable 2 Insert the end of the cable that has the cable retainer connec...

Page 13: ...et the signal generator 3 Check for error messages If error messages occur refer to Checking Signal Generator Functionality on page 6 to clear the error messages If no error messages are present the s...

Page 14: ...m up for five minutes 2 Press the front panel PRESET button to reset the signal generator 3 Check for error messages If error messages occur refer to Checking Signal Generator Functionality on page 6...