26
Operation and Service Guide
Service
Figure 4
Removing Power Sensor Shell
Disassembly Procedure
Disassemble the power sensor by performing the following steps:
CAUTION:
Disassemble the power sensor only in a static free workstation. Electrostatic
discharge will render the power sensor inoperative.
1
At the rear of the power sensor, insert the blade of a screwdriver between the plas-
tic shells (Figure 4). To prevent damage to the plastic shells use a screwdriver
blade as wide as the slot between the two shells.
2
Pry alternately at both sides of the connector J1 until the plastic shells are apart.
Remove the shells and the magnetic shields.
Reassembly Procedure
1
Replace the magnetic shields and the plastic shells as shown in Figure 3. Snap the
plastic shells together.