![Agilent Technologies E4412A Operation And Service Manual Download Page 22](http://html1.mh-extra.com/html/agilent-technologies/e4412a/e4412a_operation-and-service-manual_2867103022.webp)
20
Operation and Service Guide
Performance Test
Table 5
Power Sensor SWR and Reflection Coef ficient for the E4413A
a
Frequency
System Rho
Uncertainty
Actual
Measurement
Maximum
Rho
50 MHz to <100 MHz
±
0.010
0.095
100 MHz to <8 GHz
±
0.010
0.087
8 GHz to <18 GHz
±
0.010
0.095
18 GHz to 26.5 GHz
±
0.015
0.115
a.Specifications apply to instruments with serial prefix US3848 and above. For