![Agilent Technologies E4400-60218 Installation Note Download Page 5](http://html1.mh-extra.com/html/agilent-technologies/e4400-60218/e4400-60218_installation-note_2867100005.webp)
Installation Note E4400-90344
5
Disassemble the Instrument - External Shell
1. Remove the two strap handles (item 1) from each side of the signal generator by loosening the two
screws (item 2) and each handle. Refer to Figure 1, "Instrument Cover and Associated Parts".
2. Remove the four bottom feet (item 3).
NOTE
Save all hardware so it can be reinstalled.
3. Remove the rear feet (item 4) from the signal generator by removing the four screws (item 5) that
secure them.
4. Slide the instrument cover (item 6) off the back of the signal generator.
5. Remove the cover-top chassis by removing the 11 screws that secure it.
Figure 1
Instrument Cover and Associated Parts