Installation Note E4400-90344

3

ESG Series Signal Generators: Part Number E4400-60218

Introduction

This kit enables you to replace the small fan (B1). You will perform the following steps:

• check instrument functionality

• disassemble the instrument - external shell

• remove the existing small fan (B1) assembly

• install a new small fan (B1) assembly

• verify fan operation

• reassemble the instrument

• verify instrument operation

Installation Kit Parts List

Product Affected: . . . . . . . . . . . . . . . . . . . . . . . . . .

E4400A/B, E4420A/B, E4421A/B, E4422A/B, E4423B, E4424B, E4425B,
E4426B, E4430A/B, E4431A/B, E4432A/B, and E4433A/B, E4434B,
E4435B, E4436B, & E4437B Family RF Synthesized Signal Generators

Serial Numbers: . . . . . . . . . . . . . . . . . . . . . . . . . . .

All

Options: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

All

To Be Performed By: . . . . . . . . . . . . . . . . . . . . . . .

(X) Agilent Technologies Service Center
(X) Personnel Qualified by Agilent Technologies
(X ) Customer

Estimated Installation Time: . . . . . . . . . . . . . . . . .

1 hour

Estimated Verification Time: . . . . . . . . . . . . . . . .

1  minute

Item

Quantity

Description

Part Number

1

1

B1 Fan Kit Assembly

E4400-60222

2

4

Rivet NY Flat

0361-1341

3

4

Screw 3mmx10mm

0515-0374

4

1

Installation Note

E4400-90344

Summary of Contents for E4400-60218

Page 1: ...Part Number E4400 90344 Printed in USA January 2000 Installation Note Agilent Technologies ESG Family Signal Generators Small Fan B1 Replacement Kit Kit Part Number E4400 60218...

Page 2: ...to this material including but not limited to the implied warranties of merchantability and fitness for a particular purpose Agilent Technologies shall not be liable for errors contained herein or fo...

Page 3: ...ation Installation Kit Parts List Product Affected E4400A B E4420A B E4421A B E4422A B E4423B E4424B E4425B E4426B E4430A B E4431A B E4432A B and E4433A B E4434B E4435B E4436B E4437B Family RF Synthes...

Page 4: ...he power to the signal generator The green LED should light and the instrument performs a self check 3 When the display lights check to see if the ERR annunciator is on 4 If the ERR annunciator is on...

Page 5: ...nstrument Cover and Associated Parts 2 Remove the four bottom feet item 3 NOTE Save all hardware so it can be reinstalled 3 Remove the rear feet item 4 from the signal generator by removing the four s...

Page 6: ...that attach the fan assembly B1 to the instrument chassis Use a knife blade to pry up the head of the plunger portion of the rivet Once the head of the plunger is adequately raised use a pair of long...

Page 7: ...nnect the power cord to the signal generator 2 Turn on the signal generator and verify that the fan operates Reassemble the Instrument 1 Replace the cover top chassis and secure it with the 11 screws...

Page 8: ...8 Installation Note E4400 90344...