Register-Based Programming 57
Appendix B
Selecting Channels Using Your Own Relays
Figure B-1 shows the pin-outs for connectors P200 and P400 on the driver
portion of the module (see Figure 2-3). Connector P200 has connections for
remote relay modules (RMD) 0, 1, and 3. Connector P400 has RMD 2, 4,
and 5.
Figure B-1. P200 and P400 Connector Pin-outs
Figure B-2 shows a simplified schematic of a remote relay module (such
as RMD,1 for example). P1 connector pins 1, 3, 4, and 5 set the Remote
Module Code (RMC). For the 75 Ohm relay modules, the code is FF11
h
.
These pins set the code for RMC00-RMC03, RMC10-RMC13,
RMC30-RMC33 on P200 and RMC20-RMC23, RMC40-RMC43, and
RMC50-RMC53 on P400.
RMC10
RMC12
PWR2
RMD1(10)
RMD1(8)
RMD1(7)
RMD1(5)
PWR2
RMD1(2)
RMD1(0)
RMC00
RMC02
PWR2
RMD0(10)
RMD0(8)
RMD0(7)
RMD0(5)
59
57
55
53
51
49
47
45
43
41
39
37
35
33
31
29
27
25
PWR2
23
RMD0(2)
21
RMD0(0)
19
RMC30
17
RMC32
15
PWR2
13
RMD3(10)
11
RMD3(8)
9
RMD3(7)
7
RMD3(5)
5
PWR2
3
RMD3(2)
1
RMD3(0)
60
GROUND
58
RMC11
56
RMC13
54
RMD1(11)
52
RMD1(9)
50
PWR2
48
RMD1(6)
46
RMD1(4)
44
RMD1(3)
42
RMD1(1)
40
GROUND
38
RMC01
P200
36
RMC03
34
RMD0(11)
32
RMD0(9)
30
PWR2
28
RMD0(6)
26
RMD0(4)
24
RMD0(3)
22
RMD0(1)
20
GROUND
2
RMD3(1)
4
RMD3(3)
8
RMD3(6)
6
RMD3(4)
10
PWR2
12
RMD3(9)
14
RMD3(11)
16
RMC33
18
RMC31
28
RMD5(6)
RMD5(5)
27
RMD4(10)
RMD4(0)
RMD4(2)
PWR4
RMD4(5)
RMD4(7)
RMD4(8)
PWR4
RMC42
RMC40
RMD5(0)
RMD50(2)
PWR4
13
5
3
1
9
7
11
21
17
15
19
23
25
RMD5(4)
RMD5(3)
RMD5(1)
GROUND
RMC43
RMC41
RMD4(1)
RMD4(3)
RMD4(4)
RMD4(6)
PWR4
RMD4(9)
RMD4(11)
14
6
4
2
10
8
12
22
18
16
20
24
26
RMD2(10)
RMD5(10)
RMC22
57
RMD2(0)
RMC52
RMC50
PWR4
RMD5(8)
RMD5(7)
PWR4
RMD2(8)
RMD2(7)
RMD2(5)
PWR4
RMD2(2)
41
35
31
29
33
39
37
49
45
43
47
53
51
55
RMC20
59
P400
58
RMC21
RMC23
RMD2(11)
RMD2(9)
PWR4
RMD2(6)
RMD2(4)
RMD2(3)
GROUND
RMC51
RMC53
RMD5(11)
RMD5(9)
PWR4
RMD2(1)
42
36
32
30
34
40
38
50
46
44
48
54
52
56
GROUND
60
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com