4 Miscellaneous errors and recovery techniques
Using Bravo Diagnostics to reclaim reagents from syringes
18
AssayMAP Bravo Platform Error Recovery Guide
Using Bravo Diagnostics to reclaim reagents from syringes
You can use the following workflow to recover from a
W-axis was commanded to an
invalid destination
message that occurs when the AssayMAP Bravo Platform is
conducting an aspirate or dispense task that does not involve the wash station at deck
location 1.
Workflow
Perform the steps in the following workflow in the order given.
Step
For this task...
See...
1
If necessary, turn the red button
clockwise on the emergency-stop
pendant, and remove any object
that is interrupting the Light
Curtain.
—
2
If the syringes contain reagents
that you want to recover, open
Bravo Diagnostics and reclaim
the reagents before aborting the
run.
“Using Bravo Diagnostics to reclaim
reagents from the syringes” on
page 19
3
If the syringes do not contain
reagents to be recovered, abort
the run.