background image

STEP

Dionysos

Job: 210%XXXX5511-0002

5965-5601E

Comp: 04-28-00

Time: 11:20

OP: NN

PDFINFO H 5 6 0 1 - 0 2

2

Product Specifications

Input

DC:

Maximum voltage (any terminal to any other terminal or

chassis):

120 Vdc

AC rms:

Maximum voltage (any terminal to any other terminal or

chassis):

120 V rms

Maximum current (per channel common, non-

inductive):

50 mA

Maximum power per channel:

1 VA

DC

Maximum thermal offset

per channel, differential

Hi-Lo:

µ

V

Closed channel resistance:

100 

± 10%

Insulation resistance

(between any two points):

10E9 

Insulation resistance

(Hi to Lo, power off):

n/a

AC

Minimum bandwidth

(-3 dB, 50 

source/load):

10 MHz (protection resistors shorted)

Crosstalk (channel-to-

channel):

100 kHz:

–70 dB

10 MHz:

–20 dB

Both:

n/a

Closed channel

capacitance:

<

150 pF Hi-Lo, 

<

150 pF Lo-Guard, 

<

2000 pF

Guard-Chassis

General Characteristics

Relays:

Reed relays

Break-before-make

Relays open on power down

Relays open on power up

Minimum relay life:

No load:

10E8 operations

Rated load:

10E7 operations

Screw terminal wire size:

16 to 26 AWG (1.5, 1.2, 0.9, 0.75, 0.5 mm)

Scanning rate:

600 channels/s typ.

General Specifications

VXI Characteristics

VXI device type:

Register based, A16, slave only

Size:

B

Slots:

1

Connectors:

P1

Shared memory:

None

VXI busses:

None

C-size compatibility:

Requires E1403C

Instrument Drivers

See the Agilent Technologies Website (http://www.agilent.com/find/

inst_drivers) for driver availability and downloading.
Command module

firmware:

Downloadable

Command module

firmware rev:

A.01

I-SCPI Win 3.1:

Yes

I-SCPI Series 700:

Yes

C-SCPI LynxOS:

Yes

C-SCPI Series 700:

Yes

Panel Drivers:

Yes

VXI

plug&play Win

Framework:

Yes

plug&play Win 95/NT

Framework:

Yes

VXI

plug&play HP-UX

Framework:

No

Summary of Contents for Agilent E1345A

Page 1: ...ranged into two banks of eight switches each having its own common high low and guard screw terminals accessible on the terminal block Tree switched common high low and guard screw terminals and analog bus connections provide access to all 16 channels The module may be programmed as either a single 1x16 three wire multiplexer two independent 1x8 three wire multiplexers for four wire Ω or a 1x8 six...

Page 2: ...000 pF Guard Chassis General Characteristics Relays Reed relays Break before make Relays open on power down Relays open on power up Minimum relay life No load 10E8 operations Rated load 10E7 operations Screw terminal wire size 16 to 26 AWG 1 5 1 2 0 9 0 75 0 5 mm Scanning rate 600 channels s typ General Specifications VXI Characteristics VXI device type Register based A16 slave only Size B Slots 1...

Page 3: ... P mm H2O 0 02 Air Flow liter s 0 10 Ordering Information Description Product No 16 Channel Low Offset Relay Multiplexer E1345A Service Manual E1345A 0B3 Japan Japanese Localization E1345A ABJ 3 Yr Retn to Agilent to 1 Yr OnSite Warr E1345A W01 Extra terminal block for the E1345A E1345 80001 Module Current IPM IDM 5 V 0 2 0 01 12 V 0 13 0 01 12 V 0 0 24 V 0 0 24 V 0 0 5 2 V 0 0 2 V 0 0 ...

Page 4: ...PDFINFO H 5 6 0 1 0 4 4 Signal Conditioning Components Current Shunt ...

Page 5: ...consulting services Online Internet access for Agilent product information services and support www agilent com find tmdir VXI product information www agilent com find vxi Defense Electronics Applications www agilent com find defense_ATE Agilent Technologies VXI Channel Partners www agilent com find vxichanpart Agilent Technologies HP VEE Application Website www agilent com find vee Agilent Techno...

Reviews: