-4-
To Make an Insertion Loss Measurement
1. Choose the measurement parameters or recall an
instrument state.
(or
)
2. Calibrate the analyzer.
3. Connect the
equipment.
(or
)
4. Interpret the measurement / save results.
5. Calculate insertion loss and end of cable.
Insertion Loss Calculation
= marker value
÷
2
= 7.36 dB
÷
2 = 3.68 dB
End of Cable Calculation
=
λ ÷ 2
where:
λ
(wavelength) = c
×
V
f
÷ ∆
f
∆
f = 10 MHz between bumps
V
f
(velocity factor)
≈ 1.0
c
≈
10
9
ft/sec
λ
= 10
9
ft/sec
÷
10
7
cycles/sec = 100 ft
= 100 ft
÷
2 = 50 ft
PRESET
MEAS 1
Reflection
S11
AVG
System Bandwidth Narrow 250 Hz
CAL
1-Port
User 1-Port
MARKER
900
MHz
SAVE RECALL
Define Save Data ON
Prior Menu Save State