Chapter 4
179
Assembly Replacement
Procedure 1. Spectrum Analyzer Cover
Procedure 1. Spectrum Analyzer Cover
Removal/Replacement
1. Disconnect the line-power cord, remove any adapters from the front
panel connectors, and place the spectrum analyzer on its front panel.
2. If an 85620A Mass Memory Module or 85629B Test and Adjustment
Module is mounted on the rear panel, remove it. Loosen (but do not
remove) the four rear-bumper screws, using a 4 mm hex wrench.
Pull the cover assembly off towards the rear of the instrument.
CAUTION
When replacing the spectrum analyzer cover, use caution to avoid
damaging any cables.
3. When installing the cover assembly, be sure to locate the cover air
vent holes on the bottom side of the spectrum analyzer. Attach with
the four screws loosened in step 2, and tighten the four screws
gradually to ensure that the cover is seated in the front frame gasket
groove.
4. Torque each screw to 40 to 50 inch-pounds to ensure proper gasket
compression to minimize EMI.
Figure 4-1
Hinged Assemblies
NOTE
shows an 8560 E-series instrument. In the assembly removal
and replacement procedures the words “left” and “right ” assume you
are facing the front panel of the instrument, as shown in
with A14 and A15 to your left, and A2 through A5 on your right. The
8560 EC-series instrument is identical except the A2 board is smaller.
Summary of Contents for 8564EC
Page 17: ...25 1 General Information ...
Page 37: ...47 2 Adjustment Diagnostic Software ...
Page 77: ...89 3 Manual Adjustment Procedures ...
Page 129: ...161 3a Manual Adjustment Procedures 3335A Source not Available ...
Page 142: ...175 4 Assembly Replacement ...
Page 194: ...Chapter 4 257 Assembly Replacement Procedure 13 A21 OCXO Figure 4 34 A21 OCXO Mounting Screws ...
Page 196: ...259 5 Replaceable Parts ...
Page 218: ......
Page 219: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ...303 6 Major Assembly and Cable Locations ...
Page 234: ...315 7 General Troubleshooting ...
Page 238: ...Chapter 7 319 General Troubleshooting Introduction Figure 7 2 Ribbon Cable Connections 1 of 3 ...
Page 239: ...320 Chapter7 General Troubleshooting Introduction Figure 7 3 Ribbon Cable Connections 2 of 3 ...
Page 242: ...Chapter 7 323 General Troubleshooting Introduction Figure 7 5 Service Cal Data Menu ...
Page 271: ...352 Chapter7 General Troubleshooting Block Diagram Description Figure 7 6 Functional Sections ...
Page 283: ...364 Chapter7 General Troubleshooting Block Diagram Description ...
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ...377 8 ADC Interface Section ...
Page 291: ...Chapter 8 381 ADC Interface Section Introduction Figure 8 2 A3 Test Connector Pin Locations ...
Page 321: ......
Page 322: ...427 9 IF Section ...
Page 356: ...Chapter 9 461 IF Section A5 IF Assembly Figure 9 13 Detailed IF Adjust Signature 5 ...
Page 376: ......
Page 377: ......
Page 378: ......
Page 379: ...487 10 Controller Section ...
Page 394: ...521 11 Synthesizer Section ...
Page 451: ......
Page 452: ......
Page 453: ......
Page 454: ...593 12 RF Section ...
Page 489: ...628 Chapter12 RF Section A15 RF Assembly Figure 12 10 10 MHz TTL Reference at U304 Pin 13 ...
Page 491: ......
Page 492: ...633 13 Display Power Supply Section ...
Page 504: ......
Page 505: ...671 14 Component Level Information Packets ...
Page 507: ......