85103F Performance Upgrade Installation Manual
19
The 85103F Upgrade Package (8510A/B to 8510C)
Checking the 8510C
Checking the 8510C
85101C Functional Check
Does the analyzer display any internal self-test failure on the self-test status LEDs or
display? If not, the instrument passed the internal self tests, and functions fully. If a
self-test fails, or a running error is displayed, refer to the Agilent Technologies 8510C
On-Site Service Manual.
85102 Functional Check
Use these procedures to check that the LO phase lock-out and trigger-in are functioning
properly.
1. Turn both the 8510 and the source off (three line switches).
2. On the 85102:
a. From the rear panel, remove the test-set interface cable.
b. Plug the service adapter (supplied with the upgrade package) into J1 Test Set
Interconnect (the multi-pin rear-panel connector).
c. Connect the BNC end of the service adapter to the rear-panel 20 MHz output.
d. Connect the ANALOG ± 10V BNC to the SWEEP IN 0-10V BNC (failure to do this
causes an error).
3. Turn on the 8510, and ignore the system bus address error.
4. Press
AUXILIARY MENUS SYSTEM
.
5. Select
[MORE]
,
[SERVICE FUNCTIONS]
,
[TEST MENU]
.
6. Press
2
,
2
,
=MARKER
.
7. Press
2
,
=MARKER
.
8. Press
=MARKER
.
9. Run all of the tests: Press
M/u
(hexadecimal B),
=MARKER
.
10. After the tests pass, return to the 8510 Log Mag display:
Press
=MARKER
.
Press
.
(hexadecimal F),
=MARKER
.
Press
.
(hexadecimal F),
=MARKER
again.
Press
1
,
5
,
=MARKER
.
11. At this point, disregard any running error messages and continue with