3- 2
85051B
Use, Maintenance, and Care of the Devices
Electrostatic Discharge
Electrostatic Discharge
Protection against electrostatic discharge (ESD) is essential while connecting, inspecting, or cleaning
connectors attached to a static-sensitive circuit (such as those found in test sets).
Static electricity can build up on your body and can easily damage sensitive internal circuit elements when
discharged. Static discharges too small to be felt can cause permanent damage. Devices such as calibration
components and devices under test (DUT), can also carry an electrostatic charge. To prevent damage to the
test set, components, and devices:
•
always
wear a grounded wrist strap having a 1 M
resistor in series with it when handling components
and devices or when making connections to the test set.
•
always
use a grounded, conductive table mat while making connections.
•
always
wear a heel strap when working in an area with a conductive floor. If you are uncertain about the
conductivity of your floor, wear a heel strap.
•
always
ground the center conductor of a test cable before making a connection to the analyzer test port
or other static-sensitive device. This can be done as follows:
1. Connect a short (from your calibration kit) to one end of the cable to short the center conductor to the
outer conductor.
2. While wearing a grounded wrist strap, grasp the outer shell of the cable connector.
3. Connect the other end of the cable to the test port.
4. Remove the short from the cable.
shows a typical ESD protection setup using a grounded mat an wrist strap. Refer to
for information on ordering supplies for ESD protection.
Figure 3-1
ESD Protection Setup
Summary of Contents for 85051B
Page 4: ...iv 85051B ...
Page 7: ...85051B 1 1 1 General Information ...
Page 12: ...1 6 85051B General Information Clarifying the Terminology of a Connector Interface ...
Page 13: ...85051B 2 1 2 Specifications ...
Page 19: ...85051B 3 1 3 Use Maintenance and Care of the Devices ...
Page 37: ...85051B 4 1 4 Performance Verification ...
Page 40: ...4 4 85051B Performance Verification Recertification ...
Page 41: ...85051B 5 1 5 Troubleshooting ...
Page 43: ...85051B 5 3 Troubleshooting Troubleshooting Process Figure 5 1 Troubleshooting Flowchart ...
Page 46: ...5 6 85051B Troubleshooting Contacting Agilent ...
Page 47: ...85051B 6 1 6 Replaceable Parts ...
Page 50: ...6 4 85051B Replaceable Parts Introduction Figure 6 1 Replaceable Parts ...