4-2
85032F
Performance Verification
Introduction
Introduction
The performance of your calibration kit can only be verified by returning the kit to Agilent
Technologies for recertification. The equipment and calibration standards required to
verify the specifications of the devices inside the kit have been specially manufactured and
are not commercially available.
How Agilent Verifies the Devices in This Kit
Agilent verifies the specifications of these devices as follows:
1. The residual microwave error terms of the test system are verified with precision
airlines and shorts that are directly traced to NIST (National Institute of Standards
and Technology). The airline and short characteristics are developed from mechanical
measurements. The mechanical measurements and material properties are carefully
modeled to give very accurate electrical representation. The mechanical measurements
are then traced back to NIST through various plug and ring gages and other mechanical
measurements.
2. Each calibration device is electrically tested on this system. For the initial (before sale)
testing of the calibration devices, Agilent includes the test measurement uncertainty as
a guardband to guarantee each device meets the published specification. For
recertifications (after sale), no guardband is used and the measured data is compared
directly with the specification to determine the pass/fail status. The measurement
uncertainty for each device is, however, recorded in the calibration report that
accompanies recertified kits.
These two steps establish a traceable link to NIST for Agilent to the extent allowed by the
institute’s calibration facility. The specifications data provided for the devices in this kit
are traceable to NIST through Agilent Technologies.
Summary of Contents for 85032F
Page 4: ...iv 85032F ...
Page 7: ...1 1 1 General Information ...
Page 12: ...1 6 85032F General Information Preventive Maintenance ...
Page 13: ...2 1 2 Specifications ...
Page 18: ...2 6 85032F Specifications Electrical Specifications ...
Page 19: ...3 1 3 Use Maintenance and Care of the Devices ...
Page 33: ...4 1 4 Performance Verification ...
Page 36: ...4 4 85032F Performance Verification Recertification ...
Page 37: ...5 1 5 Troubleshooting ...
Page 41: ...6 1 6 Replaceable Parts ...
Page 45: ...A 1 A Standard Definitions ...