5-35
Servicing
Performance Tests
Configuring the pattern generator
This procedure shows you, step-by-step, how to configure the Agilent 70841B
pattern generator settings for the performance tests and adjustment proce-
dures. It is assumed, that the Agilent 70311A clock source is configured in the
modular measurement system (MMS) mainframe as a slave to the pattern gen-
erator module. This allows you to change the clock source’s frequency from
the pattern generator. If your clock source is not configured as a slave module
to the pattern generator, you can still set its frequency using its own menu.
Consult the user manuals for these products to learn how to correctly install
and operate them.
1
Turn on the MMS system, and display the menu for the pattern generator.
2
Press the green
INSTR PRESET
key.
3
Press the left-side
select pattern
softkey.
4
Press the
2
23
–
1
softkey.
5
Press the left-side
trg o/p clk o/p
softkey.
6
Press
TRIGGER PAT CLK
so that
CLK
is underlined.
This sets the trigger mode.
7
Press
CLOCK FREQ
, and enter the clock frequency.
8
Press
CLOCK AMPLTD
, and enter the clock amplitude.
9
Press the left-side
dat o/p err-add
softkey,
more 2 of 2
, and then
DAT TRM 0V –2V
, so
that
–2V
is underlined.
This sets the data termination level.
10
Press
more 2 of 2
and then
DATA ECL
.
Summary of Contents for 83430A
Page 2: ...Agilent 83430A Lightwave Transmitter User s Guide...
Page 9: ......
Page 11: ......
Page 16: ...1 5 Getting Started Step 1 Inspect the Shipment...
Page 25: ......
Page 31: ...2 6 Making Measurements Using the Agilent 83430A...
Page 44: ...3 Specifications 3 3 Regulatory Information 3 6 Specifications and Regulatory Information...
Page 50: ...3 7 Specifications and Regulatory Information Regulatory Information...
Page 51: ......
Page 63: ...5 6 Servicing General Information...
Page 69: ...5 12 Servicing Troubleshooting...
Page 97: ...5 40 Servicing Adjustment Procedures Figure 5 11 Location of A3 power supply adjustments...
Page 103: ......