Performance Tests
Performance Tests
84
Agilent 81618A/9A and Agilent 81622B/3B/4B/6B/7B/8B User’s Guide, Second Edition
Performance Test for the Agilent 81628B
Page 3 of 3
Model Agilent 81628B Optical Head
Report No.________
Date ________________
Test
No.
Test Description
Min.
Spec.
Result Max.
Spec.
Measurement
Uncertainty
I
Linearity Test (without sphere)
For Calculations you may want to use the appropriate sheet
Range
P
DUT
[dBm]
P
DUT
[dBm]
Loss [%]
+40
+9
________
_______
<± 0.46 %
+40
+3
________
_______
<± 0.46 %
+30
+3
________
_______
<± 0.46 %
+30
-7
________
_______
<± 0.46 %
+20
-7
________
_______
<± 0.46 %
+20
-17
________
_______
<± 0.46 %
+10*
-17*
Reference
0.0
<± 0.46 %
+10
-27
________
_______
<± 0.46 %
0
-27
________
_______
<± 0.46 %
0
-37
________
_______
<± 0.46 %
-10
-37
________
_______
<± 0.46 %
-10
-47
________
_______
<± 0.49 %
-20
-47
________
_______
<± 0.49 %
-20
-57
________
_______
<± 0.71 %
II
Accuracy Test (with sphere)
[
µ
W]
measured at _______ nm
(1310nm) Output Power
9.64
µ
W
_______
10.36
µ
W
measured at _______ nm
(1550nm) Output Power
9.64
µ
W
_______
10.36
µ
W
III
Noise Test
[pW]
_______
< 5 nW