Performance Tests
Performance Tests
44
Agilent 81618A/9A and Agilent 81622B/3B/4B/6B/7B/8B User’s Guide, Second Edition
3
Move to the Laser Source channel, move to the wavelength
parameter, [
λ
], press
Enter
, select the lower wavelength source, and
press
Enter
.
4
If you are using an Agilent 81657A Laser Source make sure you
initialize the Agilent 8156A Optical Attenuator with 30 dB
attenuation.
5
Turn the instruments on, enable the laser source and allow the
instruments to warm up for at least 20 minutes.
6
Perform the following sub-procedure for both Power Meters:
a
Move to the Power Meter channel.
b
Move to the wavelength parameter, [
λ
], press
Enter
, enter the
wavelength of the laser source, and press
Enter
.
c
Move to the calibration parameter, [
CAL
], press
Enter
, set the
calibration parameter to zero, and press
Enter
.
d
Move to the averaging time parameter, [
AvgTime
], press
Enter
,
move to <
500 ms
>, and press
Enter
.
e
Move to the power parameter, [
P
], press [Pwr unit], move to
<
Watt
>, and press
Enter
.
7
Make sure the optical input of the Device Under Test (DUT),
8162xA/B, is not receiving any light by placing a plastic cap over the
input. Move to the DUT Power Meter channel, press [Menu], move to
<
Zero
>, press
Enter
.
8
Ensure that the Agilent 8156A output is disabled. Move to the
reference Power Meter channel that uses the Agilent 8162xB Optical
Head and Agilent 81618A Optical Head Interface module, press
[Menu], move to <
Zero
>, press
Enter
.
9
Enable the Agilent 8156A output and change the attenuation until
the reference Power Meter displays 10.00 mW.
10
Connect the attenuator output cable to the DUT Power Meter. Note
the power value returned from the DUT, [
P
], from the display and
note the result in the test record.
11
Move to the Laser Source channel, move to the wavelength
parameter, [
λ]
, press
Enter,
select the longer wavelength source,
and press
Enter
.
12
Repeat steps 4 to 10 at the second wavelength with the
corresponding source.