Setting Up Generic and Advanced Signals
Setting Up a Burst Signal
46
Agilent 81133A/81134A Pulse Generator User’s Guide, August 2002
Setting Up a Burst Signal
Task
Set up a burst signal consisting of a 32-bit data pattern repeated twice
at a period of 500 ps with 50 ps delay and 100 ps pulse width.
The data pattern is:
10100000000000000000000000000000
The amplitude is 2.0 V and the offset is 0 V.
Start the output at the rising edge of an external signal with 1 V
threshold applied at the start input.
Using the Graphical User Interface
Instrument Settings
To set the instrument parameters:
1
Disable the outputs.
2
In the instrument panel, choose the
Burst
mode.
3
Set the burst repetition to
2
.
4
Set the
Period
to
500 ps
.
Arming Source
Triggered by:
External Signal
-1 V
Generated
Signal
1 V
Delay = 50 ps
Width = 100 ps
Period = 500 ps