Instruction Guide
Replacing the NPD Bead
3
Remove the 3 T- 10 Torx screws from the bead assembly.
4
Gently lift up and remove the old bead assembly. Avoid
bumping the bead on the sides of the collector.
5
Remove the protective cap covering the new bead.
Bead assembly
N O T E
For a new Blos bead, the bead should be centered as shown in
you notice that the bead assembly does not appear like the assembly
shown in
, carefully center the bead. Be sure to not contaminate
the bead during this process.