CI Maintenance
To Clean the CI Source
5977B Series MSD Operating Manual
241
To Clean the CI Source
Materials needed
•
Abrasive paper (5061-5896)
•
Alumina abrasive powder (8660-0791)
•
Aluminum foil, clean
•
Cloths, clean (05980-60051)
•
Cotton swabs (5080-5400)
•
Glass beakers, 500 mL
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Solvents
•
Acetone, reagent grade
•
Methanol, reagent grade
•
Methylene chloride, reagent grade
•
Ultrasonic bath
Preparation
1
Disassemble the CI source. (See
“To Disassemble the CI Source”
2
Collect the following parts to be cleaned for a CI source: (See
•
Repeller
•
Source body
•
Drawout plate
•
Drawout cylinder
•
Ion focus lens
•
Entrance lens
These are the parts that contact the sample or ion beam. The other parts
normally should not require cleaning.