Chapter 7 Analyzing, Transferring, and Printing Measurement Data
Setting Up the Data Analysis Graph
Getting Started Guide
7-11
7
Figure 7-5. Machine Information screen
Creating a compensation table
To create a compensation table to correct the machine tool for linear
positioning errors, select Comp Table from the Show Data Set screen. The
Linear Compensation Table screen is displayed (Figure 7-6).
Compensation is based on the data as it appears in the Show Data Set
screen. The Linear Compensation Table screen presents compensation
values in the user’s machine units and can interpolate where target
positions do not match the machine’s compensatable positions.
Compensation can be based on the mean values for forward and reverse
travel separately or on the combined mean.
Summary of Contents for 5530
Page 1: ...Agilent 5530 Dynamic Calibrator Getting Started Guide ...
Page 2: ......
Page 3: ...Agilent 5530 Dynamic Calibrator Getting Started Guide ...
Page 10: ...Contents viii Getting Started Guide ...
Page 13: ...1 Introducing the Agilent 5530 ...
Page 20: ...Chapter 1 Introducing the Agilent 5530 Types of Measurements 1 8 Getting Started Guide 1 ...
Page 21: ...2 Making Laser Measurements Safely ...
Page 26: ...Chapter 2 Making Laser Measurements Safely Protecting the Optics 2 6 Getting Started Guide 2 ...
Page 27: ...3 Installing the Software and Connecting Components ...
Page 45: ...4 Using the Agilent 5530 for the First Time ...
Page 66: ......
Page 67: ...5 Setting Up the Software for a Measurement ...
Page 91: ...6 Making the Measurement ...
Page 101: ...7 Analyzing Transferring and Printing Measurement Data ...
Page 113: ...8 Ensuring Repeatability Accuracy and Resolution ...
Page 127: ...9 Troubleshooting and Maintenance ...
Page 133: ...Glossary ...
Page 138: ...Glossary G 6 Getting Started Guide Glossary ...
Page 144: ...Index Index 6 Getting Started Guide ...
Page 149: ......
Page 150: ... 10747 90061 Manual Part Number 10747 90061 Printed in U S A OCTOBER 13 2008 ...