4
Performance Tests
4
Turn on the frequency counter, set counter to:
•
Function: Freq
•
Termination: 50
Ω
: AC
•
Auto Trig: Off
•
Trigger level: 25 mV
•
Slope: Pos
•
Sensitivity: Hi
5
When the
READY
LED stops flashing and is stable, observe the frequency
displayed on the frequency counter.
6
Record the date measured and actual reading on line 2 of the
Wavelength test
After replacement of the laser tube and power supply a wavelength test
should be performed. Wavelength tests require specialized and expensive
equipment. Wavelength tests can be performed at a qualified service center
and the factory. Contact your local Agilent service facility for a quote. If
you want to explore doing the wavelength test on your own, please consult
with your local bureau of standards. You can search the National Institute
of Standards and Technology in the USA at the following web site:
www.nist.gov.
Summary of Contents for 5517B
Page 1: ...Agilent Technologies Agilent 5517B BL C D DL FL Laser Head User s Guide...
Page 10: ...x Agilent 5517B BL C D DL FL User s Guide Contents...
Page 12: ...xii Agilent 5577B BL C D DL FL User s Guide Figures...
Page 14: ...xiv Agilent 5577B BL C D DL FL User s Guide Tables...
Page 24: ...10 Agilent 5517B BL C D DL FL User s Guide 1 General Information...
Page 28: ...14 Agilent 5517B BL C D DL FL User s Guide 2 Laser Head at a Glance...
Page 52: ...38 Agilent 5517B BL C D DL FL User s Guide 4 Performance Tests...
Page 64: ...50 Agilent 5517B BL C D DL FL User s Guide 5 Adjustments...
Page 86: ...72 Agilent 5517B BL C D DL FL User s Guide 7 Service...
Page 94: ...80 Agilent 5517B BL C D DL FL User s Guide 8 Specifications...
Page 98: ......
Page 99: ......
Page 101: ......
Page 102: ...05517 90077 Manual Part Number 05517 90077 Printed in U S A JUNE 18 2009...