Program
Installation
The
adjustment
program
must
be
installed
in
the
computer
using
the
following
procedures
.
Making
a
W
orking
Copy
Copy
the
content
of
the
adjustment
program
disk
to
a
working
disk
or
to
the
harddisk
to
prevent
the
program
le
from
accidental
deletion
or
destruction.
Use
the
working
disk
or
harddisk
and
store
the
original
disk
in
a
safe
place
.
Program
Execution
P
erform
the
following
steps
to
start
the
adjustment
program.
Note
Install
the
program
prior
to
execution.
Refer
to
\Program
Installation
".
1.
Turn
on
the
computer
and
bring
up
the
B
ASIC
system.
Refer
to
the
T
able
1-2
for
the
required
language
extensions
.
2.
Set
the
Mass
Storage
Unit
Specier
to
the
drive/directory
where
the
adjustment
program
exists
using
the
MSI
command.
3.
Load
the
program,
\ADJ4338B "
by
pressing
LOAD
"ADJ4338B"
4
Return
5
(or
4
Enter
5 ).
4.
Execute
the
program
by
pressing
RUN .
Note
In
the
program,
the
4338B
is
referred
as
\UUT".
Note
If
the
adjustment
being
performed
follows
right
after
EEPROM
replacement,
change
switch
A2SW1-1
to
\CLOSE",
so
that
the
4338B
will
ignore
the
self
test
errors
at
power-on.
(Return
the
switch
to
\OPEN"
when
the
adjustment
has
been
completed.)
Refer
to
Figure
B-1
for
the
switch
location.
2-4