Chapter 10
183
Command Reference
:INIT
1
0
. C
o
mm
and
Refe
re
nc
e
:INIT
Syntax
:INITiate[:IMMediate]
Description
Starts the trigger system once.
If the trigger system is in the idle state, executing this command starts the trigger system
and, after measurement is executed once, returns the instrument to the idle state.
If the trigger system is not in the idle state or successive startup of the trigger system is
enabled (set to ON with the
command), executing this command causes an
error and the command is ignored.
For details on the trigger system, refer to “Trigger system” on page 52. (No query)
Related commands
Equivalent key
sequence
No equivalent key is available on the front panel.
:INIT:CONT
Syntax
:INITiate:CONTinuous {ON|OFF|1|0}
:INITiate:CONTinuous?
Description
Determines whether to start up the trigger system successively.
For details on the trigger system, refer to “Trigger system” on page 52.
Parameters
This setup is initialized to ON when using the
command and to OFF when
command.
Query response
{1|0}<newline><^END>
Related commands
Equivalent key
sequence
No equivalent key is available on the front panel.
Description
ON or 1 (initial value)
Enables successive startup.
OFF or 0
Disables successive startup.
Summary of Contents for 4288A
Page 18: ...16 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Page 212: ...210 AppendixA Manual Changes Manual Changes ...
Page 248: ...246 AppendixB Information for Replacing 4278A with 4288A Comparison of Interfaces ...
Page 258: ...256 AppendixC Status Reporting System Using the Status Reporting System ...
Page 288: ...286 Index Index ...