Agilent 4155B/4156B VXIplug&play Driver User’s Guide, Edition 4
2-41
Driver Functions
hp4156b_stress
hp4156b_stress
This function forces the stress defined by the hp4156b_setStress,
hp4156b_addStressSyncIv, and hp4156b_addStressSyncPulse functions.
Syntax
ViStatus _VI_FUNC hp4156b_stress(ViSession vi, ViPInt32 status);
Parameters
vi
Instrument handle returned from hp4156b_init( ).
status
Stress output status. 0 (no error), or 1 to 255 (error status).
Remarks
The following parameters must be set within the range shown in the following table.
The period is a parameter of hp4156b_setStress function. And width and delay are
parameters of hp4156b_addStressSyncPulse function.
period
width
delay
2E-6 to 100E-6 sec
1E-6 to 99.9E-6 sec
0 to 100E-6 sec
100E-6 to 1E-3 sec
1E-6 to 999E-6 sec
0 to 1E-3 sec
1E-3 to 10E-3 sec
10E-6 to 9.99E-3 sec
0 to 10E-3 sec
10E-3 to 100E-3 sec
100E-6 to 99.9E-3 sec
0 to 100E-3 sec
100E-3 to 1.0 sec
1E-3 to 999E-3 sec
0 to 1.0 sec
1.0 to 10.0 sec
10E-3 to 9.99 sec
0 to 10.0 sec