Triggering
See also “Triggering,” starting on page 80 in chapter 3.
The meter’s triggering system allows you to generate triggers either
manually or automatically, take multiple readings per trigger, and insert
a delay before each reading. Normally, the meter will take one reading
each time it receives a trigger, but you can specify multiple readings (up
to 50,000) per trigger. You can also set an infinite trigger.
Triggering the meter from the remote interface is a multi-step process
that offers triggering flexibility.
•
First, you must configure the meter for the measurement by
selecting the function, range, resolution, etc.
•
Then, you must specify the source from which the meter will accept
the trigger. The meter will accept a software (bus) trigger from the
remote interface, a hardware trigger from the rear-panel
Ext Trig
(external trigger) terminal, or an immediate internal trigger.
•
Then, you must make sure that the meter is ready to accept a trigger
from the specified trigger source (this is called the
wait-for-trigger
state).
The diagram on the next page shows the meter’s triggering system.
First time SCPI users,
see page 183
4
Chapter 4 Remote Interface Reference
Triggering
151
Summary of Contents for 34420A
Page 1: ......
Page 2: ......
Page 13: ...1 Quick Start ...
Page 31: ...2 Front Panel Operation ...
Page 55: ...3 Features and Functions ...
Page 117: ...4 Remote Interface Reference ...
Page 200: ...5 Error Messages ...
Page 213: ...6 Application Programs ...
Page 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Page 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Page 247: ...7 Measurement Tutorial ...
Page 274: ...8 Specifications ...
Page 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Page 294: ......