4
Specifications
Measurement Accuracy
Agilent 4339B test conditions*:
1. Warm up time:
≥
30 minutes
2. Ambient temperature: 23 °C ± 5 °C
3. Test cable length:
≤
1.5 meters
4. Open error correction performed
5. Long measurement time setting
6. Contact check: off
: Measured resistance value in W
amperes
Vs
: Source voltage in volts
Vo
: 0.1 V@Vs
≤
200 V, 0.5 V@Vs > 200 V
Accuracy example:
To determine the accuracy of a meas-
urement use Figure 1, “Conversion
diagram”.
For example: determine the accuracy
of a 5 M
Ω
(= 5 x 10
6
Ω
) measurement
at 50 Vdc.
Rm
= 5 x 10
6
Ω
Vs
= 50 V
The intersection of Rm running paral-
lel to the 10
6
Ω
diagonal line intersects
the vertical Vs line at the second row
from the bottom of the diagram. Mov-
ing horizontally across to Table 1, the
following equation is found:
100 Vo + (2.5 x 10
-7
x
Rm
)
0.5 +
Vs
Entering the values for Rm, Vo, and
Vs yields an accuracy of ±0.725%.
Table 1. Agilent 4339B measurement accuracy (±% of reading)
Resistance RM (
Ω
)
Current IM (A)
Resistance measurement
Accuracy
Current measurement
Accuracy