Site Requirements for the 2100 Bioanalyzer Instrument
2
Physical Specifications of the 2100 Bioanalyzer Instrument
Physical Specifications of the 2100 Bioanalyzer Instrument
on page 71 lists the physical specifications of the Agilent 2100
Bioanalyzer instrument.
Table 1
Physical specifications
Type
Specification
Comment
Weight
10 kg (22 lbs)
Dimensions
162 — 412 — 290 mm
(6.4 — 16.2 — 11.4 in)
Width — depth — height
Line voltage
100-240 VAC ±10%
Wide-ranging capability
Line frequency
50-60 Hz ±5%
Power consumption
30 W / 60 VA
Ambient operating temperature
15-27 °C (59-80 °F)
The assays set-up to be analyzed on a chip
temperature of 30 °C. If the ambient
temperature is above 30 °C the assays are
analyzed using that temperature.
Ambient non-operating
temperature
-40-70 °C (-40-158 °F)
Humidity
<70%, at 15-27 °C (59-81 °F)
Non-condensing
Operating altitude
Up to 2000 m (6500 ft)
Non-operating altitude
Up to 4600 m (14950 ft)
For storage of the instrument.
Safety standards:
IEC, EN,CSA, UL
Installation Category II, Pollution
Degree 2
For indoor use only.
ISM Classification
ISM Group I Class B
According to CISPR 11
Summary of Contents for 2100 Bioanalyzer System
Page 1: ...Agilent Technologies Agilent 2100 Bioanalyzer System Installation and Safety Guide ...
Page 14: ...14 1 Safety Information Legal Notice ...
Page 24: ...24 1 Sicherheitsinformationen Rechtlicher Hinweis ...
Page 34: ...34 1 Informations relatives à la sécurité Avis légal ...
Page 44: ...44 1 Información sobre seguridad Aviso legal ...
Page 54: ...54 1 安全に関する情報 法的規制 ...
Page 56: ...56 1 安全信息 注意 仪器安全 本仪器是一类安全仪器 有保护接地端 且已根据国际安全标准进 行制造和测试 ...
Page 64: ...64 1 安全信息 法律公告 ...
Page 76: ...76 3 Unpacking the 2100 Bioanalyzer System Damaged Packaging ...
Page 86: ...86 5 Starting the 2100 Bioanalyzer Instrument Preparing the Assay ...
Page 88: ...88 6 Maintenance of the 2100 Bioanalyzer Instrument Preparing the Assay ...
Page 91: ......