![Agilent Technologies 16753A Service Manual Download Page 87](http://html.mh-extra.com/html/agilent-technologies/16753a/16753a_service-manual_2867833087.webp)
86
Chapter 3: Testing Logic Analyzer Performance
Performance Test Record
Performance Test Record
LOGIC ANALYZER MODEL NO. (circle one): 16753A 16754A 16755A 16756A
Logic Analyzer Serial No.
Work Order No.
Date:
Recommended Test Interval - 2 Year/4000 hours
Recommended next testing:
TEST EQUIPMENT USED
Pulse Generator Model No.
Oscilloscope Model No.
Pulse Generator Serial No.
Oscilloscope Serial No.
Pulse Generator Calibration Due Date:
Oscilloscope Calibration Due Date:
MEASUREMENT UNCERTAINTY
Clock Rate
Pulse Width (Eye Width)
Pulse Generator Frequency Accuracy:
8133A: 1% of setting
Oscilloscope Horizontal Accuracy:
54845B: ±[((0.007%) (
∆
t)+(full scale/(2x memory depth))+30
ps]
≅
30 ps
Oscilloscope Display Resolution: 54845B: ±5 ps
Setting: 150 MHz + 2% = 153 MHz
300 MHz + 2% = 306 MHz
Pulse Width setting: 930 ps
PWmax(worst case): 930 ps+30 ps+5 ps = 965 ps
TEST RESULTS
Logic Analysis System Self-Tests
(Pass/Fail):
Performance Test: Minimum Master to Master Clock Time and Minimum Pulse Width
300 Mb/s mode
600 Mb/s mode
Pulse Generator Settings
Freq: 150 MHz plus test margin
Pulse Width: 1 ns less test margin
Freq: 300 MHz plus test margin
Pulse Width: 1 ns less test margin
Test Criteria
Test 1 of 2
Eye Finder locates an
eye for each bit
Test 2 of 2
Correct number of
occurrences
detected
Test 1 of 2
Eye Finder locates an
eye for each bit
Test 2 of 2
Correct number of
occurrences
detected
Pod 1 Results
(pass/fail)
:
Pod 2 Results
(pass/fail)
:
Pod 3 Results
(pass/fail)
:
Pod 4 Results
(pass/fail)
:
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com