Chapter 3
21
Operation
Performing Fixture Compensation
Figure 3-1
Performing short compensation
3. Fasten the electrode securing screws.
NOTE
Do not fasten the screws too tightly to prevent the electrodes from damaged.
4. Follow the instruction manual that came with your instrument to perform a
measurement to obtain short compensation data.
Performing Load Compensation
Generally, there is no need to perform load compensation. If you have any
standard device or you need to keep consistency in measured data, perform load
compensation.
The load compensation procedure is as follows.
1. Fasten the electrode securing screws at the HIGH and LOW sides with load
leads caught in the electrodes.
NOTE
Do not fasten the screws too tightly to prevent the electrodes from damaged.
2. Follow the instruction manual that came with your instrument to perform a
measurement to obtain load compensation data.