![Agilent Technologies 11982A Operation And Service Manual Download Page 26](http://html1.mh-extra.com/html/agilent-technologies/11982a/11982a_operation-and-service-manual_2868241026.webp)
1-19
General Information
Electrostatic Discharge Information
Electrostatic Discharge Information
Electrostatic discharge (ESD) can damage or destroy electronic components.
All work on electronic assemblies should be performed at a static-safe work
station.
shows an example of a static-safe work station using two
types of ESD protection:
• Conductive table-mat and wrist-strap combination.
• Conductive floor-mat and heel-strap combination.
Both types, when used together, provide a significant level of ESD protection.
Of the two, only the table-mat and wrist-strap combination provides adequate
ESD protection when used alone.
To ensure user safety, the static-safe accessories must provide at least 1 M
Ω
of
isolation from ground. Refer to
for information on
ordering static-safe accessories.
W A R N I N G
These techniques for a static-safe work station should not be used
when working on circuitry with a voltage potential greater than
500 volts.