Installation and Preparation for Use
Preparing the Agilent 11982A for Use
Connecting the Agilent 11982A to a Power Source
The lightwave converter is a portable instrument and requires no physical
installation other than connection to a power source.
C A U T I O N
Do not
connect ac power until you have verified that the line voltage is correct,
the proper fuse is installed, and the line voltage selector switch is properly
positioned, as described in the following paragraphs. Damage to the equipment
could result.
Power Requirements
Setting the Line Voltage Selector
Use the following procedure to set the Agilent 11982A’s voltage selector to the
voltage range (100, 120, 220, or 240 V) corresponding to the available ac volt-
age.
C A U T I O N
Before connecting the lightwave converter to the power source, you must set
the rear-panel voltage selector correctly to adapt the lightwave converter to the
power source. An improper selector setting can damage the lightwave
converter when it is turned on.
1
Pry open the fuse holder door with a small screwdriver.
Table 2-3. Agilent 11982A Power Requirements
Characteristic
Requirement
Input Voltage
100, 120, 220, or 240 V (±10%)
Frequency
47 to 63 Hz
Power
75 VA (maximum)