![Agilent Technologies 11970 Series User Manual Download Page 57](http://html.mh-extra.com/html/agilent-technologies/11970-series/11970-series_user-manual_2867527057.webp)
Chapter 3
57
Performance Tests
Conversion Loss and Frequency Response
P
e
rf
orm
a
nce
T
est
s
Conversion Loss and Frequency Response
Specifications
Conversion Loss:
For a CW RF input power of less than -20 dBm
11970K: 24 dB maximum
11970A: 26 dB maximum
11970Q: 28 dB maximum
11970U: 28 dB maximum
11970V: 40 dB maximum
11970W: 47 dB maximum
Frequency Response:
For an LO amplitude between 14.5 and 16.0 dBm
11970K: ±1.9 dB
11970A: ±1.9 dB
11970Q: ±1.9 dB
11970U: ±1.9 dB
11970V: ±2.1 dB
11970W: ±3.0 dB
For an LO amplitude between 14.0 and 18.0 dBm
11970K: ±2.8 dB
11970A: ±2.8 dB
11970Q: ±2.8 dB
1197OU: ±2.8 dB
11970V: ±2.8 dB
11970W: ±4.0 dB