Chapter 3
53
Performance Tests
Performance Test Procedures
P
e
rf
orm
a
nce
T
est
s
Table 3-3. Recommended Test Equipment for 11970Q
Instrument
Critical Specifications
Recommended Model
Spectrum Analyzer
LO and IF ranges compatible with mixer
8566B
Synthesized Sweeper
Frequency: 11.0 to 16.7 GHz
Output Level: >0 dBm
8340A
Amplifier
Output Level: >18 dBm leveled
Frequency Range: 4 to 6 GHz
11975A
Amplifier
Frequency Range: 11.0 to 16.7 GHz
Output Level >15 dBm
8349A
Power Meter
Compatible with Power Sensor
436A
Power Sensor
SWR: <2.0
Q8486A
Power Sensor
SWR: <1.3 @ 6 GHz
8485A
Directional Coupler
1
1. Calibration data for the coupling ration between the output arm and the auxil-
iary arm is necessary for accurate measurements.
Coupling: 20 dB
Directivity: >30 dB
Auxiliary Arm SWR: <1.2
Q752D
Isolator
Insertion Loss: <2 dB
Isolation: >20 dB
SWR: <1.5
Q365A
Frequency Tripler
Conversion Loss: <20 dB
Spacekom Microwave
2
TKa-1
2. Honeywell, Inc. Spacekom Microwave Center, Santa Barbara, CA
Adapter
SMA female to Type N female
P/N 1250-1404
Adapter
SMA female to SMA female
P/N 1250-1158
Adapter (2 required)
Type N male to SMA female
P/N 1250-1250
Cables (4 required)
Connectors: SMA male
P/N 5061-5458
Cable
Connectors: SMA male
Loss: <1.0 dB @ 20 GHz
P/N 8120-4396
Cable
Connectors: BNC male
11086A