8
1159A 1GHz Active Differential Probe
Specifications and Characteristics
Specifications and Characteristics
Specifications
Range
Input Configuration
Ground Connector
True Differential (+ and inputs), with shield
Input Coupling
DC AC coupling obtained by installing an AC coupling
adapter
Gain Accuracy at 1 kHz
2%
Maximum Input Voltage
Either input from ground
< ±42 V
CMRR
at 70 Hz: 80dB
at 1 MHz: 40dB
at 100 MHz: 25dB
at 500 MHz: 19dB
at 1 GHz: 13dB
Ð
!
Mode
No attenuator
÷10 attenuator
÷20 attenuator
Differential
< ±400 mV
< ±4 V
< ± 8V
Common
< ±16 V
< ±42 V
< ±42 V
Offset (Common) < ±1.6 V
< ±16 V
< ±32 V