![Agilent Technologies 1159A User Manual Download Page 37](http://html1.mh-extra.com/html/agilent-technologies/1159a/1159a_user-manual_2867089037.webp)
36
1159A 1GHz Active Differential Probe
Calibration Test Record
Calibration Test Record
Agilent Technologies
1159A 1 GHz Differential Probe
Serial No.:_______________________
Certification Date:_________________
Tested By:_______________________
Recommended Test Interval: 1 Year
Recommended Date of Next Certification:_________
Certification Temperature:_____________________
_______________________________
_______________________________
Test
Limit Minimum
Results
Limit Maximum
Accuracy at 1 kHz into 50
input of 54845 A
Min Gain 98%
Max Gain 102%
Differential Mode Range
Att 1; Gain 1
Att 10; Gain 10
No Clipping
Pass
_____
_____
Fail
_____
_____
N/A
CMRR at 70 Hz
80 dB
N/A
CMRR at 1 MHz
40 dB
N/A
CMRR at 100 MHz
25 dB
N/A
CMRR at 500 MHz
19 dB
N/A
CMRR at 1 GHz
13 dB
N/A
Ω