4 Printer
settings
The purpose of this section is to show additional settings of the printer that are
not covered by the Printer Installation Wizard.
The following table lists these settings.
Printer Setting
Importance
Accessible via ...* Details see...
Film calibration
optional
Key-operator menu
Default densitometer
optional
Key-operator menu
Film view
optional
Key-operator menu
Beep on empty tray
optional
Key-operator menu
RF Tag
optional
Service menu
Site data
mandatory
Service menu
4.1 Film
calibration
Accessible via menu:
Key-operator menu - 2 Change settings - 4 Image
Quality - 1 General
Purpose of the setting:
A) Define whether automatic calibration shall be
executed at insertion of a new film pack
B) If set on ON, define the number of filmpacks,
after which the film calibration shall be
executed.
Default value:
ON - every 5 filmpacks
The film calibration menu is accessible via key-operator menu on the keypad
and key-operator menu via browser.