![AGFA Drystar 5302 Technical Documentation Manual Download Page 48](http://html.mh-extra.com/html/agfa/drystar-5302/drystar-5302_technical-documentation-manual_2866645048.webp)
Controls, Connections and Set Up Procedures
DD+DIS150.05E
Printed copies of this document are not controlled.
The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Chapter 1 / 14
Drystar 5302
Edition 1, Revision 0
4.6 Site
data
Accessible via menu:
Browser pages - Service Engineer tools - Enter
key-operator command session - 2 Change
settings - 1 General
Purpose:
Fill in site data to have an entry in the infocounter
about the installation site.
Default value:
None
The site data menu is accessible via the browser pages only
Note: It is mandatory to fill in site data. Otherwise after each switch on /
reboot of the printer a hint is displayed to enter the site data.
Example browser
screen:
536601fq2.cdr
Date:
Time:
Language
Keypad contrast (60 ... 80) Typical 68
Keypad backlight (0 ... 100)
* Site name
* Site address
* City
* Country code (tel)
* Telephone
Modem
* Department
* Contact person
* Delivery address
E-mail address
Comments
Key-Operator - Change general settings
19
08
30
00
English
Chinese Simpl.
Japanese
5 beeps
English
68
Beep on empty tray
50
Oct.
2004
Accept changes Discard changes Reset Values
Welcome to this
DryStar 5302 printer
Figure 9
Agfa Company Confidential
(Typ
5366/100)