DD+DIS150.05E
Repair and Service
Adjustments and Calibrations
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
3
Checks
Following checks can be performed at the printer:
Adjustment
To be done:
Details see:
Density reproduction check
During maintenance. Refer to
chapter 9, Maintenance.
During troubleshooting
procedure
Check for dust stripes / TH
cleaning procedure
During troubleshooting
procedure
Resistor check
During troubleshooting, to check
whether print stripes are caused
by the thermal head.
3.1
Density Reproduction Check
REQUIRED TOOLS:
Service PC with Density reproduction check files, available via MedNet, GSO Library
path: Hardcopy => General Hardcopy Info => General Services Info => Application
Info => Density Reproduction Check Excel Sheet (
/
)
Purpose:
With the density reproduction check the difference for a specific film type between
optical density
defined
in the sensitometric calibration file and the
printed
optical
density, measured with an external densitometer is checked.
This is a plausibility check for the CDM: In case the measured density values are out
of the specified range, the CDM requires a re-calibration.
This procedure must be carried out with every used film type:
(1) In the browser pages select menu "Service Engineer Tools - Enter Service
Command Session - 3 Print Image - 1 Test Image".
(2) Print out LinCheck.zip and measure the density of the eleven 100 % – 0% fields
with a calibrated MacBeth TR-924, X-Rite 331, 341, X-Rite 301, X-Rite 310 or
Gretag D200-2 densitometer.
(3) To check linearity fill in measured values in the the density reproduction check
Edition 1, Revision 1
DRYSTAR 5302
Chapter 3.6 / 31
05-2009 Type
5366/100
Agfa Company Confidential