Supported sizes
15x30 to 43x35
in portrait and landscape orien-
tation
Lifetime
Expected lifetime for the bucky
10 years
Environmental conditions
Table 7: Environmental conditions for the X-ray system
Environmental Conditions (during storage and transport)
Temperature (ambient)
between -15° and 50° Celsius
Humidity (non condensing)
between 15 and 90 % relative humidity
Atmospheric pressure
between 70 and 106 kPa
Environmental Conditions (during normal operation)
Temperature (ambient)
between 10° and 35° Celsius
Humidity (non condensing)
between 30 and 75 % relative humidity
Atmospheric pressure
between 70 and 106 kPa
Maximum altitude
3000 m
For overall system environmental conditions, the environmental conditions of
the DR Detector or image plate should be taken into account. Refer to the
related User Manual for environmental conditions for the DR Detector or
image plate. When using the DR Detector or image plate inside the bucky, take
into account that the temperature inside the bucky can be up to 5°C higher
than the temperature in the X-ray room.
Bucky | Product Information |
53
3253F EN 20210312 1125