Estimated product life (if regular-
ly serviced and maintained ac-
cording to Agfa instructions)
100000 RAD
Pixel Matrix
Pixel size
139 µm (H,V)
Pixel matrix
3072 (H) x 3072 (V)
Active pixel matrix
3052 (H) x 3052 (V)
Fill factor
100 %
Detector type
Amorphous Silicon
Active area size
42,7 cm (H) x 42,7 cm (V)
Partnumber 7965
Partnumber 7964
Maximum Linear Dose us-
ing RQA5
50 µGy
75 µGy
Minimum Modulation Transfer Function (MTF) using RQA5
1 lp/mm
0.45
0.45
2 lp/mm
0.15
0.15
3 lp/mm
0.05
0.05
Typical Detective Quantum Efficiency (DQE) using RQA5
(at 2.1 µGy dose level) (at 4.0 µGy dose level)
0 lp/mm
0.59
0.25
1 lp/mm
0.41
0.20
2 lp/mm
0.23
0.10
3 lp/mm
0.11
0.03
3.6 lp/mm
0.06
0.01
Minimum Signal Noise Ratio for 1mR
SNR
115:1
115:1
Conversion screen
CsI
GOS
56