Wrong modality position selected
Details
The active modality position on the X-ray system does
not match the selected modality position on the NX
Workstation.
Cause
The modality position has been modified on the gener-
ator console.
This situation applies only to Siemens generators.
Brief Solution
To use another modality position for a scheduled expo-
sure:
1.
On the NX workstation, click
Edit
in the Image De-
tail Pane and select an exposure type for the correct
modality position.
2.
Continue the exposure workflow.
DR Generator Sync Box, DR Retrofit Solution | Problem solving |
49