Quality Control
In order to establish and maintain consistent image quality, a regular
evaluation of image quality is advised.
The printer contains an automatic QC feature that has been designed to
comply with the grayscale reproduction constancy test, according to the
international standard IEC 1223-2-4.
Local Regulations may require other procedures.
The QC procedure consists of two main steps:
• Before initial use, establishing a number of reference values that will be
used for further follow-up and verifying initial image quality.
• After establishing these values, performing regular daily, weekly and
annual quality tests.
Note
:
Repeat the two main steps of the Hard Copy quality
procedures for each input tray loaded with general radiography
films.
The results of these tests are recorded on Quality Control Charts.
The QC image has several additional fields where the QC data can be filled in.
This image should be filed as part of the QC procedure.
Related Links
Establishing the reference values and verifying image quality
on page 94
Performing quality control (QC) tests
on page 102
on page 95
on page 124
Topics:
•
Establishing the reference values and verifying image quality
•
•
Establishing the daily operating reference density levels
•
Establishing the image geometry reference values
•
Verifying Acceptable Spatial Resolution, Artifact Levels and Low
Contrast Visibility
•
Performing quality control (QC) tests
Crystal Imager | Advanced Operation (key-operator mode) |
93
3831B EN 20210601 1655
Summary of Contents for 5366/300
Page 1: ...Crystal Imager 5366 300 User Manual 3831B EN 20210601 1655 ...
Page 12: ...12 Crystal Imager Crystal Imager 3831B EN 20210601 1655 ...
Page 57: ... The display Crystal Imager Introduction 57 3831B EN 20210601 1655 ...
Page 141: ...Crystal Imager Plug Play Installation Manual 141 3831B EN 20210601 1655 ...