Summary of Contents for 200

Page 1: ...rinted in Germany 01 2002 Order No DD DIS013 02E 1UT44X1 1 Piece UT44X MA1 ADC Solo Type 5155 100 200 2nd Edition ADC Solo Digitizer Type 5155 100 200 Base VME This documentation is separately available Order No DD DIS013 02E ...

Page 2: ...performed by qualified technical personnel Note Electrical repairs and connections must only be performed by a qualified electrician Mechanical repairs and connections must only be performed by a qualified technician CE Declaration The CE Declaration CE Conformity becomes invalid if the product is changed without explicit consent of the manufacturer This applies to all parts not only to safety ele...

Page 3: ...ep 4 3 1 Diagnostics 4 3 1 1 Questioning of the customer 4 3 1 2 Infocounter Analysis 4 3 1 3 How to evaluate the Infocounter 4 3 1 4 Clear Infocounter 5 3 1 5 Visual Check 5 3 2 Safety Switches 6 3 3 Inside 6 3 4 Cassette Unit 6 3 5 VME Rack 7 3 6 Scan Unit 8 3 7 Transport Unit 8 3 8 Erasure Unit 9 3 9 Cassettes 11 3 10 Image Plates 11 4 Checking the Image Quality 12 4 1 Test Cycles 12 4 2 Exposu...

Page 4: ...Maintenance DD DIS013 02E Section 12 II ADC Solo Edition 1 Revision 10 Type 5155 100 200 ...

Page 5: ...ds you must consult the respective manufacturers to avoid danger for operating personnel or the environment Do not override or disconnect the integrated safety features Switch off the ADC Solo before performing any maintenance work or repairs Disconnect the ADC Solo from the mains before making repairs or performing any maintenance activities As is the case for all technical devices the ADC Solo m...

Page 6: ...ted in the column details During the maintenance procedure always consider the safety instructions see TECHNICAL DOCUMENTATION section 1 1 Please check if it is necessary to include country specific regulations as additional maintenance points Only for Systems with DRA Contract In systems with DRA Contract the infocounters are checked and evaluated in regular intervals by the GSC If there is an in...

Page 7: ...999 0895 0 Vacuum cleaner CM 9 9999 0896 0 Dirt bags for vacuum cleaner 10 x ABC Code EFOJH ADC Cleaner Commercially available Dust brush Commercially available Lint free cloth Commercially available Soft dust brush 2 5 Required Spare Parts The following assortment represents a complete copy of the Maintenance assortment CM 051550100733 according to the R I M L assortment categorization in the spa...

Page 8: ... zip xxxx stands for the serial number 6 Start an editor e g notepad or wordpad 7 Open the file D_ infocounter 0 infocounter txt 8 Evaluate the infocounter file MG3000 Infocounter Figure 1 3 1 3 How to evaluate the Infocounter Evaluation of infocounters txt What to check in the infocounter Comment 1 1 Device Info Serial number and Installation date Compare device serial number to section 11 Techni...

Page 9: ...or amplitude 60 are exceeded check the last images on the workstation thoroughly for jitter Change galvo only if necessary 5 3 Retries Many retries 1 have to be investigated They usually lead to less throughput of the device Compare it with frequent error codes 5 7 Error History Check the last occurred errors in between two maintenance how often they appeared as well as the CBF cycles between fail...

Page 10: ...essed The power supply is only interrupted if the front door interlock A is opened A B Figure 2 3 3 Inside Vacuum cleaner Lint free cloth 1 Vacuum the inside of the digitizer and wipe it 3 4 Cassette Unit Screw driver Dust brush Roller CM 9 5145 9100 0 Toothed belt CM 9 5145 5195 0 Opener mechanism 1 Remove panel on right hand side of the digitizer 2 Clean the opener mechanism with a soft dust bru...

Page 11: ...pull the roller off its shaft In case of recorded cassette feeding problems replace the roller anyway and in any case replace it once a year B A Figure 4 Belt 5 Exchange the transport belt once a year 3 5 VME Rack Vacuum cleaner VME fan General 1 Check the fan function of the VME rack fan by hand There is no air filter to be replaced 2 Remove visible dust and dirt deposits by means of a vacuum cle...

Page 12: ... rollers have to be cleaned in place and must not be removed To move the scan rollers just turn the drive of the slow scan motor manually C Discharge brush 3 Check the discharge brush A for visible wear and heavy dirt clean by hand or replace B if necessary B A A C Figure 5 3 7 Transport Unit Soft cloth ADC Cleaner if not available use water Suction cups CM 9 5145 6550 0 Filter 1 Pull the hoses of...

Page 13: ...f vacuum problems check the suction cups by bending up the edges check for tears and replace if necessary A C B Figure 7 3 8 Erasure Unit Soft cloth ADC Cleaner if not available use water 10 x 100 W lamps CM 9 0450 6553 0 1 Open the front door 2 Pull the erasure unit out and remove it Lamps 3 Remove pane A 4 Pull all lamps carefully out of their sockets B 5 Check all parts of the erasure unit for ...

Page 14: ...ADC Cleaner for cleaning all the surfaces with the exception of the inner side of the large glass plate This side must not be cleaned with anything wet since a gelatin layer is attached to it 7 Check the KG2 filter A for damage and replace if necessary A B Figure 9 Fan 8 Insert the new lamps carefully Do not touch the glass bulbs with your bare fingers Use a soft cloth to insert the lamps 9 Check ...

Page 15: ... is noticed check further cassettes 2 Check the following items of the cassette Outside condition Hinges Locking Opening leaf springs Aluminum label ALU LABEL 514609AA CDR 3 Attach missing aluminum label The digitizer needs the aluminum label to recognize ADC cassettes 3 10 Image Plates Visual Check 1 Check if there are scratches on the surface 2 Check if edges are loose as an indication for mecha...

Page 16: ...luate all the images on the Processing Station and a printer Following check the flat field for homogenous field or stripes criteria The hard disk of the digitizer provides two flat field samples for quality comparison Repeat this procedure for all formats on site Flatfield 1 Print the flat field sample provided by the digitizer Start the service program Select from the service menu Checks Send fl...

Page 17: ...ice that all exposure parameters are approximate values Rotating Anode Cathode x ray tube Figure 10 3 Identify the cassette on the ID Station In the Patient name field type a name and a cassette format e g Flatfield 18 x 24 In the First name field type the serial number of the digitizer e g SN1356 In the Birth date field type the current date e g 20012001 use date format xxyyzzzz for day month yea...

Page 18: ...S013 02E Section 12 14 ADC Solo Edition 1 Revision 10 Type 5155 100 200 4 Insert the cassette into the digitizer and print the image on a printer with a window setting of 0 6 without changing the level setting Figure 11 ...

Page 19: ...wing sketches Calibration lines Blurred dark lines in slow scan direction on the flat field see beside Expose another flat field and compare it again with the sample If there are still unacceptable effects you have to redo shading calibration as described in section 6 6 Expose another flat field and compare it again with the sample If there are still unacceptable effects please contact the Support...

Page 20: ...he scan brush to remove it Expose another flat field and compare it again with the sample If there are still unacceptable effects please contact the Support Center Figure 14 5 Completion of Maintenance 1 Confirm the maintenance by signing the checklist 2 Make a backup of the system on floppy 3 Inform the customer about what was done during the maintenance and which repairs need to be done in next ...

Page 21: ......

Page 22: ...e stampato in Germania Editado por AGFA and the Agfa Rhombus are trademarks of Agfa Gevaert AG Germany Herausgegeben von Published by Edité par Pubblicato da Editado por Agfa Gevaert AG Fototechnik Tegernseer Landstraße 161 D 81539 München ...