HYPERGLYCAEMIC WARNING ALARM:
Flashing
message with the time. Test result flashes, with
double beep (if volume is ON).
MEANING:
If the Hyperglycaemic Warning Alarm is set, your
glucose test result is higher than the pre-set threshold.
KETONE MESSAGE:
Flashing message, with double
beep (if volume is ON).
MEANING:
The result of the glucose test is higher than 13.3
mmol/L.
ACTIONS:
You should check ketones. Follow your healthcare
professional’s recommendations.
TEMPERATURE MESSAGE:
Flashing message, with
double beep (if volume is ON).
MEANING:
The test has been performed outside the operat-
ing temperature range.
ACTIONS:
Move to an area with an ambient temperature
of 10 °C to 40 °C (50 °F to 104 °F). Wait for the meter and test
strips to reach the new temperature (usually 10-20 minutes)
and retest.
LOW BATTERY MESSAGE:
Flashing message
MEANING:
Battery #2 in your meter is low.
ACTIONS:
If the low battery symbol appears, battery #2 must
be replaced. Battery numbers are indicated by the numbers
printed on the pull tabs.