Control solution checks
Out of range test results may be due to:
•
Out of date or faulty control solution.
•
Out of date or faulty test strip.
•
Mistake in performing a test.
•
Watered-down control solution.
•
The JAZZ™ DoseCoach
®
is not working
properly.
•
The control solution test was done
outside the temperature range of
10°C to 40°C.
•
Not mixing the control solution
enough before using.
•
Not throwing away the first drop of
control solution and wiping the bottle
tip clean.
If the results from the control
solution tests continue to be
out of range, do not use JAZZ™
DoseCoach
®
to test your blood
glucose. Call customer service at
0800 093 1812.