background image

OptRx

PN 2006020–ENG Rev. H

Crop Sensor Settings

The Crop Sensor Settings screen is where you 
can adjust application rate settings specified by 
the OptRx Crop Sensor module. The VI Zero and 
VI Minimum settings are used as a buffer to 
prevent applying the wrong amount of N or 
applying it in areas where doing so would be 
ineffective.

To access the Settings screen, press the OptRx 
Crop Sensor button, 

located in the box containing the Map screen’s VI 
Bar Graph.

VI Zero

 — The threshold below which the Crop 

Sensor applies a zero rate. This setting is used to 
prevent applying N over bare ground or 

permanently damaged crops.You may enter in a value either by pressing 

, or you may enter in the current VI value 

by pressing 

.

VI Minimum

 — The threshold below which the Crop Sensor applies the minimum rate. This setting is used to apply a 

minimum amount of N on ground with damaged crops.You may enter in a value either by pressing 

, or you may 

enter in the current VI value by pressing 

.

÷

The minimum rate may be adjusted at the Crop Sensor screen, which is accessed by pressing the 

Algorithm 

Settings

 button.

Reference Strip

 — The name of the Reference Strip that you entered during the scanning procedure.

VI Reference

 — Value of a healthy crop biomass used in reference for calculating the proper N application rate

Algorithm Settings 

— Press this button to access the Sensor Setup screen, where you can adjust the inputs used by the 

N Algorithm to calculate to optimal Nitrogen rate. Without a valid Reference Strip this button will not be active. Please refer 
to the Crop Sensor Setup section below for information on how to decide on what inputs to use for the N Algorithm.

Crop Sensor Settings

VI Zero:

VI Minimum:

Set to 

Current

Reference Strip: Reference

VI Reference: 0.5000

Set to 

Current

Algorithm 

Settings

Set to 

Current

Set to 

Current

Summary of Contents for OptRx

Page 1: ...osed to direct water and sunlight for extended periods of time Because direct sun UV light can be destructive sensor should be oriented so that optics face the ground evenwhennotinuse Tomaximizethelif...

Page 2: ...ual Rate Actual Rate The Flow Sensor returns the actual rate being applied NOTE In some conditions the Actual Rate may increment slower than the Target Rate or its numeric values may vary before match...

Page 3: ...eir desired rates to apply any product to any crop type based on the vegetative index You can create up to 16 ranges and assign a desired rate to these ranges You also have the ability toselect your V...

Page 4: ...r field for 300 seconds 5 minutes Wheat North America Tillering Scan the most vigorous portion of your field for 300 seconds 5 minutes Europe Scan the portion of your field where the plants show an av...

Page 5: ...Barley the Growth Stage is used to select the most appropriate VI index NDVI or NDRE further it will be used for fine tuning and optimizing the N Algorithm Press when finished 3 Drive Reference Strip...

Page 6: ...trip by pressing the View All button located on the bottom right hand side of the Map screen s Equipment Tab The file name of the new reference strip appears in the Reference Strip Selection screen To...

Page 7: ...used to apply a minimum amount of N on ground with damaged crops You may enter in a value either by pressing or you may enter in the current VI value by pressing The minimum rate may be adjusted at t...

Page 8: ...viewed and set for each application Algorithm implication The Minimum Rate is the amount of nitrogen the operator wants applied to the healthiest areas of a field Keep in mind OptRx may apply very low...

Page 9: ...iewed and set for each application Algorithm implication The Pre Sidedress Rate is the amount of nitrogen that has been applied for the current year s crop before side dressing Keep in mind not all pr...

Page 10: ...field average and that the Minimum and Maximum Rate have an equal offset from the Planned Rate The Planned Rate should be set to the intended flat rate to apply if the OptRx N Algorithm was not used M...

Page 11: ...e exact values for these inputs are up to the grower ideally after consulting with an agronomist Ag Leader Technology does not provide agronomic advice Override Serial Logging Checking Logging Overrid...

Page 12: ...row sensor placement Checklist Sensors should be equally spaced apart Sensors should not be mounted on the ends of the booms Each sensor should be mounted so that it is centered over the top of a crop...

Page 13: ...mote sensing that displays a comparison of crop vigor and biomass of the scanned plant This value which is taken from a reference scan of a Reference Strip uses a ratio that indicates the health of th...

Page 14: ...here has been a drop in voltage to the OptRx Crop Sensors Solution Check voltage on the high current power cable to ensure voltage provided to the sensors is approximately 12 V Error Message Too Few S...

Reviews: